Model Number 101/815/060
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other text: d4: lot number, expiration date and h4: device manufacture date is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the external inflation cuff had came off twice.There was patient involvement but there was no injury.
 
Event or Problem Description
Additional information received.The lot number remains unknown after attempts with supply company.
 
Additional Manufacturer Narrative
Describe event or problem, 10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.Email is: (b)(6) one device sample was received in used condition without original package.Per visual inspection, the pilot balloon is detached from the inflation line.The complaint was confirmed, and no other analysis was performed.The root cause is attributed to manufacturing.A device history record (dhr) review could not be performed as the lot number was unknown.The issue has been escalated to address root cause investigation for detaching pilot balloon from inflation line tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUSELECT TRACHEOSTOMY TUBE
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17368453
Report Number3012307300-2023-07488
Device Sequence Number19802228
Product Code BTO
UDI-Device Identifier15019517077339
UDI-Public15019517077339
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/815/060
Device Catalogue Number101/815/060
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer08/27/2023
Initial Report FDA Received Date07/20/2023
Supplement Report FDA Received Date02/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-