• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GETINGE DISINFECTION AB 86-SERIES; DISINFECTOR, MEDICAL DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GETINGE DISINFECTION AB 86-SERIES; DISINFECTOR, MEDICAL DEVICES Back to Search Results
Model Number 8668
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 7th july, 2023 getinge became aware of an issue related to the 86-series washer disinfector, with the model name 8668.As it was stated, carts fell off the trolley.There was no injury or damage reported, however we decided to report the issue based on a potential as carts falling off the trolley could bring a hazardous situation for the operator and lead to serious injury if the situation was to reoccur.
 
Additional Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.H3 other text : device not returned to the manufacturer.
 
Additional Manufacturer Narrative
Getinge became aware of an issue related to the 86-series washer disinfector, with the model name 8668.While the event occurred, washer disinfector was cooperating with smart trolley¿s.As it was stated, carts fell off the trolley, due to missing stop pins.The affected parts were replaced and device returned to usage.Despite our best efforts, we cannot define the root cause of the issue.There was no injury or damage reported, however we decided to report the issue based on a potential as carts falling off the trolley could bring a hazardous situation for the operator and lead to serious injury if the situation was to reoccur.When the incident occurred, the product was directly involved, and out of specification.The device was not being used for treatment or diagnosis of the patient.Trend review of customer product complaints with the same issue involved on this type of devices reported within the last 5 years was performed but did not provide any signals that would warrant further scrutiny.We currently do not have any information that would warrant further action towards the device manufacturing or devices on the market, however as per our complaint handling processes will continue to monitor the customer experiences with the device for any future information.
 
Event or Problem Description
Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
86-SERIES
Common Device Name
DISINFECTOR, MEDICAL DEVICES
Manufacturer (Section D)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo
Manufacturer (Section G)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo
Manufacturer Contact
maria elovsson
ljungadalsgatan 11
vaxjo 
MDR Report Key17369417
Report Number9616031-2023-00017
Device Sequence Number17887000
Product Code MEC
UDI-Device Identifier07340153700277
UDI-Public(01)07340153700277
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8668
Device Catalogue NumberS-86682003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2023
Supplement Date Received by Manufacturer10/28/2024
Initial Report FDA Received Date07/21/2023
Supplement Report FDA Received Date10/28/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-