| Model Number |
1516-20-607 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon questioned whether there was a deficiency with the anterior tab of the poly as it did not properly seat when impacted.No surgical delay was reported.Doi: unknown.Dor: (b)(6) 2023.Affected side: unknown knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a1, a2, a3 and b5 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates that this was not a revision.The implant reported was never implanted.The surgeon attempted to implant it, but it would not seat when impacted, so he used a second implant that seated properly.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product 151620607, lot d22112446, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|