• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH PERCUTANEOUS INSERTION HANDLE FRN; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH PERCUTANEOUS INSERTION HANDLE FRN; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number SD03.033.001
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on 30-june-2023 during an unknown procedure, the impactor broke while inserting the nail.The procedure was completed successfully with no surgical delay.There was no reported adverse patient impact.No further information is available.This report is for a percutaneous insertion handle frn.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: initial reporter is a (b)(6), employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the percutaneous insertion handle frn.No evidence of breakage was observed.A broken fragment of the returned driving cap was stuck inside the mating hole.The investigation was not able to confirm the reported event.A dimensional inspection was not performed as it is not applicable to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the percutaneous insertion handle frn was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: yes.Dimensional inspection: n/a.Part number: sd03.033.001.Lot number: 846p235.Manufacturing site: haegendorf.Release to warehouse date: 14.09.2022.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUTANEOUS INSERTION HANDLE FRN
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17371125
Report Number8030965-2023-09082
Device Sequence Number19779196
Product Code HSB
UDI-Device Identifier10886982310699
UDI-Public(01)10886982310699
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSD03.033.001
Device Catalogue NumberSD03.033.001
Device Lot Number846P235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/30/2023
Supplement Date Received by Manufacturer08/11/2023
Initial Report FDA Received Date07/21/2023
Supplement Report FDA Received Date08/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DRIVING CAP/THREADED.; UNK - NAILS.
Patient SexUnknown
-
-