The device was returned to olympus for evaluation and the customer¿s allegation was confirmed.The device evaluation found the knife wire was broken.The fracture was charred and melted but the knife wire was measured, and there was no abnormality.It was confirmed that there were no problems with the length of the knife wire and wire sheathing, and that there were no defects in the actual product.Since the knife wire was broken, it was not possible to confirm the direction of the knife.No other abnormalities were confirmed that could lead to knife wire breakage.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation.
|
|
The customer reported to olympus, the knife wire of the single use 3-lumen sphincterotome v normally faces the 11 o¿clock direction, but it would only face the 1 o¿clock to 4 o¿clock directions.The issue was found during the procedure.The same problem occurred with the second replacement.When the person in charge of the facility checked the devices, one of them had a broken knife wire.The procedure was completed by replacing it with a third device.There were no reports of patient harm.This report is related to linked patient identifier (b)(6).
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see updates to h6 and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it¿s likely the event (abnormal direction of the knife wire) was caused by the knife wire breaking due to discharge during energization, and the knife wire did not point in the intended direction.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.2.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were being close to each other.3.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.4.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.The following is included in the instructions for use: ¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.¿ olympus will continue to monitor field performance for this device.
|