MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3507175BC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made. as such, the investigation will be closed. if the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date. this report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report. reason for device explant and/or reoperation: breast augmentation revision.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 53-year-old caucasian female patient underwent a breast augmentation revision surgery with implantation of a 175cc mentor memorygel breast implant prosthesis.Post-operatively, the patient underwent a breast augmentation revision surgery for an undisclosed reason.During the procedure, bilaterally ruptured breast implants were discovered.As a result, the patient underwent bilateral removal and replacement with 170cc mentor memorygel breast implants on (b)(6) 2023.There were no reported procedural delays or patient consequences due to the discovery.Refer to manufacturing report number 1645337-2023-08681 for the contralateral event.
|
| |
|
Additional Manufacturer Narrative
|
|
On august 01, 2023, the mentor analysis lab received the suspect medical device for evaluation.On august 07, 2023, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection of the device.Visual analysis of the returned sample revealed that the breast implant had an area of shell abrasion on the posterior view.In addition, a tear was noted within the shell abrasion measuring approximately 1.6 cm.The evaluation determined that the cause of the rupture is consistent with normal wear.Shell abrasion suggests in-vivo folding or creasing of the device.This may be the result of the following factors: continuous and sustained stresses to the device such as too small breast pocket and folding or wrinkling of the shell in the breast pocket.In some cases, the breast implants may also wear out over time.Manufacturer¿s reference number: (b)(4).
|
| |
|