• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 1012535-60
Medical Device Problem Codes Difficult to Remove (1528); Material Separation (1562); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the superficial femoral artery.Resistance was met during advancement of the absolute pro vascular stent system through the sheath, and the shaft separated in two pieces.Therefore they attempted to remove the device, but it became stuck in the sheath.The absolute pro and sheath were removed together with no issues.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported material separation was able to be confirmed.The reported difficult to advance was unable to be replicated in a testing environment due to the condition of the returned device.The reported difficult to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during advancement interactions with the sheath resulted in the reported difficult to advance.Manipulation of the device resulted in the noted device damages (wrinkled/torn sheath) and ultimately resulted in the reported shaft separation/noted jacket and inner member separations; thus resulting in the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17374598
Report Number2024168-2023-07852
Device Sequence Number14409731
Product Code NIP
UDI-Device Identifier08717648176074
UDI-Public08717648176074
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model Number1012535-60
Device Catalogue Number1012535-60
Device Lot Number2111561
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/29/2023
Supplement Date Received by Manufacturer08/10/2023
Initial Report FDA Received Date07/21/2023
Supplement Report FDA Received Date08/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
Patient Weight103 KG
-
-