The customer reported to olympus that an electrical failure occurred and no image during the hysteroscopic myomectomy and the same problem occurred after replacing all the combined devices causing a 30-minute delay in surgery and 30 minutes extended anesthesia time.There were no error messages.The procedure was completed using other urological products.The event did not affect the diagnostic or treatment outcome.There were no reports of patient harm.This event requires 12 reports.Related patient identifiers for the first combination of devices used are as follows: (b)(6) (wa00014a / hf-cable, bipolar), (b)(6) (wa00014a / hf-cable, bipolar), (b)(6) (a42011a / resection sheath, 8 mm, for 8.5 mm/26 fr.Outer sheath, abs), (b)(6)(a42021a / outer sheath, 8.5 mm / 26 fr., 2 stopcocks, rotatable), (b)(6) (wa22366a /working element, active, for resection in saline), (b)(6)(wb50402w / foot switch, double, for, esg-400).Related patient identifiers for the second combination of devices used: (b)(6) (wb50402w / foot switch, double, for, esg-400), (b)(6) (wa22366a / working element, active, for resection in saline), (b)(6) (a42011a / resection sheath, 8 mm, for 8.5 mm/26 fr.Outer sheath, abs), (b)(6) (a42021a /outer sheath, 8.5 mm / 26 fr., 2 stopcocks, rotatable), (b)(6) (wa00014a / hf-cable, bipolar), (b)(6) (wa00014a /hf-cable, bipolar).This medwatch report is for patient identifier: (b)(6).
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the findings are as follows: no abnormalities were detected with the returned device.No electrical problems could be confirmed.The output was normal.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the cause of the reported issue could not be determined as no fault or malfunction has been identified with the affected devices.The device was assessed as standard.However, device users occasionally report that the power output is too low or that the surgical result is insufficient.Inadequate surgical performance can basically be traced back to insufficient power output into the tissue.Possible causes are as follows: 1.Defective cable connection to the connected instrument -> no or only intermittent power output.2.Defective hand instrument -> no or only intermittent power output.3.Contamination of the hand instrument due to adhering tissue deposits -> reduced power output.4.Contact resistances at the contact points of the hand instrument -> reduced power output, possibly burns to the patient.5.In case of a monopolar application an incorrectly attached neutral electrode and/or a defective supply line to the neutral electrode -> reduced or only intermittent power output with the risk of burns below the applied neutral electrode.6.Incorrect setting of the power on the generator by the user.7.An incomplete adjustment of the generator -> reduced power output 8.A defect in the generator -> reduced or no power output in addition, the electrical conditions in the tissue of the surgical area may also cause the problem.However, it is rare that errors of this type can be traced back to a device defect.In this respect, a user error can be assumed as the cause in most cases.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿a suitable replacement device must be provided during an application.¿ also, information has been added to h4.Olympus will continue to monitor field performance for this device.
|