• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number 1012534-100
Medical Device Problem Codes Entrapment of Device (1212); Device Damaged by Another Device (2915)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 06/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during delivery system removal inadvertent interaction with the previously implanted stent resulted in the reported entrapment of device.Interaction/manipulation of the device resulted in the reported device damaged by another device; thus resulting in the reported tissue injury as reportedly the delivery system dragged and crushed the previously implanted stent in the tibial artery causing vessel damage.The treatment appears to be related to the operational context of the procedure as reportedly cut-down surgery was performed to remove the delivery system and restore flow to the artery.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the superficial femoral artery (sfa) with moderate calcification.The 6.0x100mm absolute pro self-expanding stent system (sess) was advanced to the target lesion and the stent was implanted without issue; however, during removal the delivery system became stuck with a stent in the tibial artery that had been implanted in a previous procedure.The delivery system dragged and crushed the implanted stent in the tibial artery causing vessel damage and the vessel had to be clamped-off.During, multiple attempts to remove the delivery system of the sess the wire access was lost and the 3-4 cm of the delivery system remained in the posterior tibial artery and the remainder of the delivery system was outside the patient's foot.Later the same day, cut-down surgery was performed to remove the delivery system of the sess and restore flow to artery.There was no adverse sequela.The additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17375183
Report Number2024168-2023-07862
Device Sequence Number19802151
Product Code NIP
UDI-Device Identifier08717648176012
UDI-Public08717648176012
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number1012534-100
Device Catalogue Number1012534-100
Device Lot Number2022462
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/28/2023
Initial Report FDA Received Date07/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
Patient SexMale
Patient Weight74 KG
-
-