| Model Number |
ORT300 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The imris customer service engineer addressed the issue on-site.A fitting securing a pressurized hydraulic line connection was identified as the source of the leak and was adjusted and tightened.The cse measured the hydraulic fluid level in the operating room table and confirmed it was within specification.The table was functionally tested and no functional issues were present or previously reported by the customer.The cse inspected the table on a follow-up site visit and confirmed the absence of any fluid leaks.A follow-up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
During routine preventive maintenance, an imris customer service engineer observed drops of hydraulic fluid outside of the hydraulic lines of the ort300 operating room table.The source of the leak was traced to a fitting on a pressurized hydraulic line.No table functional issues were identified during subsequent testing.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The imris service engineer replaced fittings and hoses within the hydraulic system that controls trendelenburg position articulation.Hydraulic fluid levels and system pressure were measured during multiple service visits and confirmed to be within specification.No table functional impact was reported by the customer or observed by the service engineer during the site visits.The service engineer performed drift testing twice during the repair period, which passed.Followup preventive maintenance has confirmed absence of further fluid leak.This report also encompasses udi related data quality updates to match brand name and udi with gudid data.
|
| |
|
Search Alerts/Recalls
|