Model Number FX431T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 06/21/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav 2.0 system (part # fx431t) was implanted during a procedure performed on an unknown date.According to the complainant, the system was believed to be clogged.The patient underwent a revision procedure on (b)(6) 2023.The complaint device will be returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Additional Manufacturer Narrative
Manufacturing and quality control data: due to the limited information and the missing product investigation is restricted to trackability of manufacturing and quality control data.The progav 2.0 shunt system was manufactured by a qualified employee in may 2022.Deviations during assembly did not occur.The product was finally tested as article fx431t and released for packaging and sterilization and was sterilized by miethke.The progav 2.0 has a normal pressure range of 0 to 20 cmh2o.The parameters after completion of the valve assembly were tested at a flow rate of 5 ml/h or 50 ml/h at set pressures of 0, 5, 10 and 20 cmh2o, and were found to meet specifications.The valve was pre-adjusted to 5 cmh2o in accordance with the specification.All tested parameters were assessed according to specifications.Based on the delivery batch, we confirm all parameters have been inspected and approved during the manufacturing.We assure that production and quality control ensure that only products that are conform to specifications are made available in the market.Conclusion information: unfortunately, due to the missing sample, we are unable to provide a meaningful analysis.An investigation was not possible because the product is not available.With reference to the reason for the complaint, we would like to point out that deposits in the valve could be responsible for the functional impairment.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 20
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key17375425
Report Number3004721439-2023-00199
Device Sequence Number3784722
Product Code JXG
UDI-Device Identifier04041906137298
UDI-Public4041906137298
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX431T
Device Catalogue NumberFX431T
Device Lot Number20059472
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2023
Initial Report FDA Received Date07/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-