SHIRAKAWA OLYMPUS CO., LTD. OES HYSTEROFIBERSCOPE Back to Search Results
Model Number HYF-XP
Medical Device Problem Code Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's reported issue of compromised image was confirmed; however, the darkening of the image was not confirmed.During the evaluation, it was determined that: forceps channel port was deformed; image guide bundle had a stain; due to damage on objective lens, a foggy image occurred; due to damage on image guide bundle, the image had a stain; due to a pinhole on channel-tube, water tightness was lost; connecting tube had a dent; due to a cut on angle wire, bending section could not be bent in down direction at all; adhesive around light guide lens was peeled; universal cord has a dent; and scratches were found on multiple components of the device.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 13 years since the subject device was manufactured.Based on the results of the investigation, the phenomenon was attributed due to stress of repeated use, external factors, or handling.The root cause of this event was unable to be identified.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer reported to olympus the oes hysterofiberscope exhibited compromised image issues and insufficient light from the light guide lens.There were no reports of patient harm or impact associated with this event.During testing and inspection of the returned device, the forceps mouthpiece deformation was revealed.This medical device report (mdr) is being submitted to capture the reportable malfunction found during device evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES HYSTEROFIBERSCOPE
Common Device Name
HYSTEROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17375930
Report Number3002808148-2023-07400
Device Sequence Number20636564
Product Code HIH
UDI-Device Identifier04953170340178
UDI-Public04953170340178
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K891451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHYF-XP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2023
Initial Report FDA Received Date07/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-