| Model Number |
2800TFX23MM |
| Medical Device Problem Codes |
Gradient Increase (1270); Difficult to Open or Close (2921); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Fatigue (1849); Syncope/Fainting (4411)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 23mm 2800tfx aortic valve underwent a valve-in-valve procedure after an implant duration of four (4) years, seven (7) months due to aortic stenosis with an approximate gradient of 50mmhg.Two leaflets were noted to be nonfunctional.The patient initially presented with dyspnea, fatigue and syncope.The tavr procedure was performed with a 23mm 9750tfx transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: updated: b4, d4, g3, g6, h2, h4, h6 the device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.The most likely cause is patient factors, including chronic kidney disease (ckd), hyperlipidemia (hld) and coronary artery disease (cad).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: updated: b4, b5, b7, g3, g6, h2, h6.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 23mm 2800tfx aortic valve underwent a valve-in-valve procedure after an implant duration of four (4) years, seven (7) months due to severe aortic stenosis with an approximate gradient of 50mmhg.Two leaflets were noted to be nonfunctional.The patient presented dyspnea, fatigue and syncope.The tavr procedure was performed with a 23mm 9750tfx transcatheter valve.
|
| |
|