| Catalog Number |
1804250-23 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6- device code 2017 clarifier: use after expirationna.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the obtuse marginal (om) coronary artery with mild calcification and mild tortuosity.The 2.50x23mm xience skypoint stent delivery system (sds) stent was deployed; however, the sds met resistance with the stent balloon during removal; therefore, a 1.5x12mm mini trek balloon was used to removed the sds.A 2.75x12mm nc trek neo balloon was used for post dilatation and the procedure was completed.There were no adverse patient sequela and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.D9 - device available for evaluation updated from ¿yes¿ to ¿no¿ h6 - medical device problem codes 2017 and 1528 removed.
|
| |
|
Search Alerts/Recalls
|