• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERENOVUS, INC. GALAXY G3 MINI 1.5MM X 3CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERENOVUS, INC. GALAXY G3 MINI 1.5MM X 3CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number GLM915030
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
The healthcare professional that during an endovascular embolization procedure, the complaint coil, a 1.50mm x 3.00cm galaxy g3 mini (glm915030 / 30889334) could not be detached and had to be removed from the patient.The detachment cable was working fine with other coils used during the procedure.There was no report of any negative patient impact.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section d.2b: procode is krd/hcg.Section e.1: the name, phone and email address of the initial reporter are not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A review of manufacturing documentation associated with this lot (30889334) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the complaint product was not returned / received.If the complaint product is returned or information provided at a later date, the file will be updated accordingly.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.A review of manufacturing documentation associated with this lot (30889334) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.With the information available and without the complaint product available to be returned for analysis, the reported issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.10.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on (b)(6) 2023.[additional information: on (b)(6) 2023, additional information was received.The information indicated that the procedure was to embolize a ruptured anterior communicating artery (acom) aneurysm.The coil was successfully removed from the patient; it was not stretched and was still attached to the delivery system when it was removed.The concomitant microcatheter used was an echelon¿ 10 microcatheter (medtronic); the same microcatheter was used to complete the case.A pre-deployment electrical check was not performed.The fault light was not seen during the case.Upon pressing the power button, all lights did not illuminate.The green system ready light did not illuminate and during the detachment cycle, the detachment light did not illuminate and the audible signal beep was not heard.All connections fit properly without application of force.The information confirmed that there was no patient injury associated with the reported issue.There was a 20-minute delay, however, whether it was clinically significant was not stated.Updated sections: b.4, e.1, e.3, g.3, g.6, h.2, h.10, and concomitant products.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product was received in the product analysis lab on 11-oct-2023.The returned product is pending evaluation.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigation finding of the returned device.The complaint device was returned for evaluation and analysis.The investigation finding is documented below.Investigation summary: a non-sterile 1.50mm x 3.00cm galaxy g3 mini was received contained in the decontamination pouch.Visual inspection was performed.It was noted that the coil was inside of the introducer.No other damages were found.The device was inspected under the microscope.Under magnification, the embolic coil was found to be in good condition (i.E., no kinked, stretched, nor with non-concentric loops); this was found still attached to the resistance heating (rh) coil.The distal outer sheath had not been softened, indicating that the detachment process was not initiated.The electrical resistance of the galaxy g3 device was measured with a multimeter, it measured 53.3 ohms o, which is within the specification range between 48.5 ohms o ¿ 56.0 ohms o.Also, the device was connected to a lab sample detachment control box (dcb), and a lab sample enpower control cable and the power was turned on.The system ready light was illuminating.Then, the detach button was pressed, and a beep sound was heard.The coil was successfully detached from the unit.The issue documented that the coil failed to detach could not be confirmed since the device met the electrical specification range.Additionally, no product defect was identified; there may have been other circumstances or issues that occurred during the use of the device that could not be replicated in the lab setting.The instructions for use (ifu) provides proper handling instructions for the device to prevent such issues from occurring.According to the risk documentation, failure to detach is a potential issue that can occur during the coil detachment due to a detachment cycle not being initiated, as a result of the device positioning unit (dpu) wire detachment zone/extended coil being positioned beyond the microcatheter tip, which can result in reported failure.A review of manufacturing documentation associated with this lot (30889334) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) does contain the following recommendation: ¿ verify that the microcoil delivery system is fully connected, and no faults are indicated on the dcb.If a fault exists, reseat all connections between the dpu, the dcb, and the connecting cable.If a fault still persists, replace the connecting cable.If this does not correct the error, replace the dcb.If the microcoil delivery system still continues to have a fault, retrieve the microcoil, re sheathing the microcoil system, and replace it with a new microcoil system.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.10.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3 MINI 1.5MM X 3CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
CERENOVUS, INC.
6303 waterford district drive
suites 215 & 315
miami FL 33126
Manufacturer (Section G)
CERENOVUS INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key17379990
Report Number3008114965-2023-00481
Device Sequence Number6220640
Product Code KRD
UDI-Device Identifier10886704080145
UDI-Public10886704080145
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K171862
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue NumberGLM915030
Device Lot Number30889334
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2023
Supplement Date Received by Manufacturer07/28/2023
08/31/2023
10/11/2023
01/26/2024
Initial Report FDA Received Date07/24/2023
Supplement Report FDA Received Date07/28/2023
08/31/2023
10/11/2023
01/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ECHELON¿ 10 MICROCATHETER (MEDTRONIC)
Patient SexUnknown
-
-