| Model Number |
17001-01 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned instrument confirmed the distal tip of the probe has sheared off.The curved probe is used to create a pilot hole in the pedicle prior to implanting a screw and as such must be malleted into the bone.The distal tip contains the smallest diameter on the instrument and is subject to the most stress during use.When used as intended the instrument would not encounter the amount of force required to fracture the tip in this manner.The instrument conformed to dimensional specifications at the time of manufacturing and passed inspection requirements with no non-conformities reported.There were no issues during the manufacture of this product that would contribute to this complaint condition.
|
| |
|
Event or Problem Description
|
|
It was reported the tip of the probe broke off during use.
|
| |
|