• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB ENVISTA SIMPLIFEYE INJECTOR; FOLDERS AND INJECTORS, INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB ENVISTA SIMPLIFEYE INJECTOR; FOLDERS AND INJECTORS, INTRAOCULAR LENS Back to Search Results
Model Number 21987
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on insertion of an intraocular lens (iol) into the patient¿s eye the iol haptic was broken.The lens was cut and the incision was enlarged to remove the lens.Another lens was successfully implanted.There was no patient impact.Additional information was requested.
 
Additional Manufacturer Narrative
The device history record (dhr) review did not find any non-conformities or anomalies related to this event.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Investigation of this event is in progress.A follow-up report will be submitted upon completion of investigation.
 
Additional Manufacturer Narrative
Additional info: d9, g3, g6, h2, h6, h10/11 the injector was not returned for evaluation.Based on the available information, the root cause of this event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVISTA SIMPLIFEYE INJECTOR
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
MEDICEL AG
dornierstrasse 11
alternhein CH-94 23
SZ   CH-9423
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key17386861
Report Number0001313525-2023-70092
Device Sequence Number12362038
Product Code MSS
UDI-Device Identifier00757770547449
UDI-Public(01)00757770547449(17)250331(10)2080D11
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K192005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number21987
Device Lot Number2080D11
Was Device Available for Evaluation? No
Date Returned to Manufacturer06/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date07/25/2023
Supplement Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ENVISTA INTRAOCULAR LENS, EYEFILL C
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-