| Catalog Number |
MB3355 |
| Medical Device Problem Codes |
Fracture (1260); Malposition of Device (2616)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Unspecified Musculoskeletal problem (4535); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a mobi-c device was removed after the patient developed post-operative neck, shoulder, and arm symptoms, kyphosis, and the implant broke at level c6/7.During the revision, the superior end plate was found to be displaced and was not well-fixed.The mobi-c device was replaced with competitive fusion devices.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 clinical sign 4581 keyword: kyphosis.Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a mobi-c device was removed after the patient developed post-operative neck, shoulder, and arm symptoms, kyphosis, and the implant broke at level c6/7.During the revision, the superior end plate was found to be displaced and was not well-fixed.The mobi-c device was replaced with competitive fusion devices.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 additional investigation type: 4110.Corrections in g1.Additional information in h6: component, investigation type, findings, and conclusions.Inspection the device was not returned, an x-ray image was provided but no conclusions could be drawn from it.Dhr review the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge¿s control.Potential root cause a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown surgical or patient factors.Device usage this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|