C.R. BARD, INC. (BASD) -3006260740 POWERPORT IMPLANTABLE PORT, SILICONE SINGLE-LUMEN, 9.6F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 1759600
Medical Device Problem Codes Fracture (1260); Material Separation (1562); Migration (4003)
Health Effect - Clinical Codes Arrhythmia (1721); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported through the litigation process that five months and seventeen days post port placement via the right subclavian vein, the port was allegedly found to be fractured.It was further reported that a piece had broken off and migrated to the patient's heart.Furthermore, attempts were made to remove the fractured piece and the rest of the remaining product but were unsuccessful in removing the defective device in its entirety due to the developing heart arrhythmias during the procedure.Reportedly, additional surgery was done to remove the fractured catheter piece.The current status of the patient is unknown.
 
Event or Problem Description
It was reported through the litigation process that five months and seventeen days post port placement via the right subclavian vein, the port was allegedly found to be fractured.It was further reported that a piece had broken off and migrated to the patient's heart.Furthermore, attempts were made to remove the fractured piece and the rest of the remaining product but were unsuccessful in removing the defective device in its entirety due to the developing heart arrhythmias during the procedure.Reportedly, additional surgery was done to remove the fractured catheter piece.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.The definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the litigation process that five months seventeen days post port placement for cancer treatment, the port allegedly fractured.It was further reported that a piece had broken off and migrated to patient heart.The patient was developed with heart arrhythmia during the removal procedure.Further an additional surgery was done to remove to fractured catheter piece.The current status of the patent is unknown.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Approximately five months and nineteen days post port placement, a ct of chest was performed which demonstrates the catheter fragment appearing to be within the right ventricle.Patient underwent ultrasound guided right common femoral vein access, a right chest port was removed and attempted foreign body retrieval procedure.On the same day the patient was placed on right new powerport clearvue slim placement with new venous access and port pocket.A referral was made to cardiothoracic surgery for evaluation of a retained catheter fragment in the right ventricle.Patient condition was stable and discharged to home.Therefore, the investigation is confirmed for the reported fracture, material separation and migration.Furthermore, the clinical condition alleged in the complaint cannot be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: a1, a3, d4 (expiration date: 05/2022), g2, g3, h6 (method) h11: b5, b7, d1, d4, h6 (result) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT IMPLANTABLE PORT, SILICONE SINGLE-LUMEN, 9.6F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key17389414
Report Number3006260740-2023-03158
Device Sequence Number19779223
Product Code LJT
UDI-Device Identifier00801741026850
UDI-Public(01)00801741026850
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K073423
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Other,Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1759600
Device Lot NumberREEV1030
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/03/2023
Supplement Date Received by Manufacturer09/20/2023
02/02/2024
Initial Report FDA Received Date07/25/2023
Supplement Report FDA Received Date09/23/2023
02/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-