• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER VALLEYLAB; ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES) Back to Search Results
Model Number E1560
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Laceration(s) (1946)
Date of Event 06/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, during loop electrosurgical excision procedure (leep), the loop disengaged from the base on one end, caused a laceration in the vagina with bleeding of 250cc.Unable to visualize as there was too much bleeding.No blood transfusion required.Procedure was aborted.Surgeon had to pack the vagina and tried again.Patient was packed twice, and also had tranexamic acid applied to cervix for hemostasis, and had a foley catheter inserted overnight.Patient was sent home and then returned to clinic the next day to be examined.There was no further bleeding when patient was assessed the following day.Patient was scheduled to have her leep in surgical daycare on (b)(6).
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALLEYLAB
Common Device Name
ELECTROCAUTERY, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17389604
Report Number1717344-2023-00919
Device Sequence Number19779244
Product Code HGI
UDI-Device Identifier10884524000237
UDI-Public10884524000237
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K931338
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE1560
Device Catalogue NumberE1560
Device Lot NumberX0123ZX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2023
Initial Report FDA Received Date07/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-