| Catalog Number |
8065752088 |
| Medical Device Problem Codes |
Noise, Audible (3273); Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that the irrigation failure occurred, and abnormal noise was heard from the roller area during cataract surgery.Irrigating solution didn't seem to be keeping up, and the anterior chamber became unstable.The surgery was completed after replacing the product with new one.There was no patient harm.Additional information received from physician confirming that the pak was replaced and the same issue was recurred, however the surgery was performed and completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.A used fluidics management system (fms) was visually inspected and no obvious defects were detected.The sample was tested using an console.The fms primed and tuned with the ultrasonic handpiece, the 0.9 millimeter aspiration bypass system tip and infusion sleeve from lab stock successfully.No abnormal sounds were heard during testing.Cassette max vacuum and irrigation flow rate were measured and met specifications.No problem was found with the returned cassette fluidics.The root cause of the customer's complaint could not be established; the returned sample functioned per specification.This complaint has been reviewed and it is determined that no further actions will be pursed at this time as the sample met specifications.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|