| Model Number |
WA29048A |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus that when using a ureteroscope, there was a broken piece.The event occurred during reprocessing.There was no patient harm associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated, and the customer¿s allegation (broken piece) due to a bent outer tube.In addition, there were dents on the window, and the optical fiber was blurry.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The cause of the reported issues is most likely due to excessive use.Olympus will continue to monitor field performance for this device.
|
| |
|