• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSIA DIABETES CARE US INC. CONTOUR NEXT ONE; BLOOD GLUCOSE METER Back to Search Results
Model Number 9763
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Missing Test Results (3267)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that her blood glucose readings were not being stored in the memory of the contour next one meter.Additionally, on some days two readings were stored in the meter's memory when only one test was performed.After the customer changed the batteries, the readings were properly stored in the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
 
Additional Manufacturer Narrative
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer weight was not provided.
 
Additional Manufacturer Narrative
The customer returned the suspected contour next one meter for evaluation.No test strips were returned.Therefore, an in-house testing was performed with the returned meter and in-house contour next test strips using blood sample, which gave a satisfactory performance.The blood glucose readings obtained with the in-house testing were properly stored in the meter's memory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR NEXT ONE
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
ASCENSIA DIABETES CARE US INC.
100 summit lake drive
valhalla NY 10595
Manufacturer (Section G)
PT PHC INDONESIA (REGISTRATION NUMBER: 3009204541)
kawasan industri mm2100 blok o
cikarang barat bekasi
jawa barat, 17520
ID   17520
Manufacturer Contact
shweta gulati
100 summit lake drive
valhalla, NY 10595
9142361830
MDR Report Key17390858
Report Number1810909-2023-00169
Device Sequence Number6834893
Product Code NBW
UDI-Device Identifier00301939763012
UDI-Public00301939763012
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9763
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/30/2023
Supplement Date Received by Manufacturer08/09/2023
Initial Report FDA Received Date07/25/2023
Supplement Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-