DEPUY IRELAND - 3015516266 ATT REV SLV LPS FEM ADP +10; KNEE FEMORAL ADAPTOR Back to Search Results
Catalog Number 151106010
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994)
Date of Event 07/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient present in office with thigh pain.Serial x-rays show lucency around stem and subsidence of sleeve leading surgeon to believe that the stem sleeve did not achieve stable ingrowth.Sleeve is found to have only achieved fibrous ingrowth and is removed.A cemented bowed stem is placed uneventfully after clearing fibrous tissue.No patient harm occurred.No surgical delay.Doi: (b)(6), 2022.Dor: (b)(6), 2023.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATT REV SLV LPS FEM ADP +10
Common Device Name
KNEE FEMORAL ADAPTOR
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17391107
Report Number1818910-2023-15148
Device Sequence Number6808800
Product Code MBH
UDI-Device Identifier10603295503415
UDI-Public10603295503415
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K202248
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number151106010
Device Lot NumberJF3696
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2023
Initial Report FDA Received Date07/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE REV LPS INSRT XXSM 16MM; ATUN FEM SLV M/L 35MM FULL POR; ATUNE PRESSFIT STR STEM16X60MM; CONTROL CABLE W/CABLE SLEEVE; CONTROL CABLE W/CABLE SLEEVE; CONTROL CABLE W/CABLE SLEEVE; CONTROL CABLE W/CABLE SLEEVE; LPS DISTAL FEM COMP XXSM RT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient Weight53 KG
-
-