| Catalog Number |
BOPT6585 |
| Medical Device Problem Codes |
No Flow (2991); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Osteolysis (2377)
|
| Date of Event |
04/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant was removed due to bone loss.Symptoms as a result of the event: abscess & inflammation site grafted: allograft.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint: (b)(4).A4: patient weight is not provided / unknown.B3: date of event is not provided / unknown.E1: initial reporter¿s title is not provided / unknown.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report both corrected and additional information to 0001038806-2023-01387.Review of the initial report confirmed that the h6: device code was reported incorrectly.It is now being reported correctly.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.Based on the investigation, no malfunction occurred upon investigation and no association between the dental implant and bone loss was found that can explain these events.Therefore, the reported event remains non-verifiable as it is a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for bone loss problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|