Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual, mechanical and electrical inspection.The analysis showed that the conductor coil was found fractured between 35 and 37 cm distal to the is4 connector pin.The fracture in this area was consistent with a crushing type of movement.Due to the location and type of damage, analysis determined it was likely caused by entrapment in the clavicle first-rib region.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|