| Catalog Number |
603508 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Per complaint (b)(4), after clinical procedure, patient experienced failure of implant to osseointegrate.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient's identifier and weight were not provided.If the requested information becomes available, a supplementary report will be submitted.The implant failure modes, failure/ loss of osseo-integration and lack of primary stability are not product related and rather are attributed to patient contraindications, conditions, or clinician error in surgical protocol.No product investigation or corrective actions required.Complaints will continue to be trended.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up submitted for additional information.Updated section b4 for report submission date, g1 for follow-up report submitter,g3 for awareness date and g6 for report type and follow-up number.Updated h2 for follow-up type.Updated section a4 & b7 for patient weight and oral hygiene.Patient identifier and other relevant history are unknown.
|
| |
|