| Catalog Number |
XIITP5413 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
04/04/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implants at tooth sites #4,6,9,12 & 13 were removed due to pain.Symptoms as a result of the event: pain & headache.Site grafted: no.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.Follow up with the customer confirmed item and lot number.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to report additional information.Zimvie received one implant for evaluation.Visual evaluation of the as returned device identified the implant with signs of use.No damage identified or signs of malfunction that would contribute to the reported event.Measurements match drawing.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are customer error (improper techniques used) and patient factors.Therefore, based on the available information, a device malfunction did not occur.The reported event was non-verifiable with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|
| |
|