| Catalog Number |
EIC7070-01 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a laryngeal cancer surgery, the electrode of the procise lw wand fell of after 15 minutes of use.The procedure was completed with non-significant surgical delay using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.Fluid is in the fluid line.The electrode has been heavily used with some erosion but no missing areas.Bio debris is present product was out of the original packaging.No packaging returned.A functional evaluation showed the opened device was tested and plugged into the controller and registered settings (7,3).The wand produced plasma as expected.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|