| Catalog Number |
730M0024 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign object was observed when the surgeon was reviewing an x-ray that was taken intra-operatively.It is believed to be the pin from a rocket reducer, but this cannot be fully determined.There are no plans to revise the patient to remove the object.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: additional method code: 4109.Corrections in d9 and h3.Additional information in h4 and h6: component, investigation type, findings, and conclusions.Device evaluation: x-rays were provided for review and used for confirming the presence of the pin post-op.The returned device matches the information in the complaint file and was examined.Visual inspection revealed that one of the pins that retains a reduction arm was missing and the device was disassembled.Root cause: root cause was unable to be determined.This event could possibly be attributed to an unknown operational event.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a foreign object was observed when the surgeon was reviewing an x-ray that was taken intra-operatively.It is believed to be the pin from a rocket reducer, but this cannot be fully determined.There are no plans to revise the patient to remove the object.
|
| |
|