| Catalog Number |
46538KA2/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The account alleges that during a fistuloplasty procedure, the black radiopaque tip of the catheter detached from the rest of the device inside the patient.The clinician attempted to remove the foreign body but was unsuccessful.No additional patient consequence to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device has been returned for evaluation.The device was visually and microscopically investigated.The complaint is confirmed.The root cause is attributed to the manufacturing process.The device history record was reviewed, and two exception documents were identified.A search of the complaint database was performed and no similar complaints for this lot number were identified.Corrective actions are in process.
|
| |
|