• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number S-45-100-120-P6
Medical Device Problem Codes Difficult or Delayed Activation (2577); Mechanical Jam (2983); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified, mildly tortuous, 70% stenosed lesion in the superficial femoral artery (sfa).The 4.5x100 supera peripheral self-expanding stent system (sess) was advanced to the target lesion and deployment was attempted, however the stent only deployed approximately a quarter of the way when significant resistance was experienced with the thumbslide.The sess was withdrawn from the anatomy and a non-abbott device was used to successfully complete the procedure.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Event or Problem Description
Subsequent to the initially filed report, additional information was received: the issue was resistance with the thumbslide.The stent was implanted in the patient.No replacement device was used.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported mechanical jam and the reported difficult or delayed activation were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the moderately calcified, mildly tortuous and 70% stenosed anatomy and/or inadvertent mishandling resulted in the noted multiple device damages (bends on the sheath sporadically, multiple bends on the shaft) thus during stent deployment resulted in the reported mechanical jam.Manipulation of the compromised device resulted in the noted bent ratchet ear (likely contributing to the reported difficulties) and ultimately resulted in the reported difficult/delayed activation.There is no indication of a product quality issue with respect to manufacture, design or labeling.H6: code 3270 was deleted, code 2577 added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17400485
Report Number2024168-2023-08070
Device Sequence Number17655902
Product Code NIP
UDI-Device Identifier08717648225987
UDI-Public08717648225987
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue NumberS-45-100-120-P6
Device Lot Number2100561
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/03/2023
Supplement Date Received by Manufacturer08/29/2023
Initial Report FDA Received Date07/26/2023
Supplement Report FDA Received Date08/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-