• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG SHUNTASSISTANT 15; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG SHUNTASSISTANT 15; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV250T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 03/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: in the first step of our investigation, we performed a visual inspection of the product.We have checked for possible damage, deformation of the housing, deposits or other abnormalities.The following observations were made during the visual inspection: no obvious defects.Permeability test: the permeability test was performed at a hydrostatic pressure difference of the pressure setting of the shuntassistant® upon receipt plus 30 cmh2o in the horizontal direction of flow.The test showed that the shuntassistant is permeable.Computer control test: to check the flow rate of the valve, the valve was tested on a miethke computer controlled testing apparatus and passed the standard test.The flow of the test liquid was reduced step by step from 60 ml/h to 5 ml/h (in accordance with iso 7197).Distilled water was used as the test-liquid.The resulting pressure was measured.The shuntassistant was tested according to standard procedure in the vertical position of the valve.At a fixed opening pressure of 15 cmh2o in the vertical position, a pressure of 15 cmh2o -2 / +4 cmh2o is expected.The results indicated that at a reference flow of 20 ml/h in the vertical position, the shuntassistant had a pressure of 15,81 cmh2o.This is within the specified tolerance of 15 cmh2o -2 / +4 cmh2o.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valve is finally opened with a special tool.After dismantling of the valve, deposits were found in shuntassistant.For more detailed visibility, the proteins/deposits in shuntassistant were coloured by using a colouring solution.[colouring solution consisting of coomassi brilliant blue r mixed with 0.6% ethanol and distilled water].Results: based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.Further actions: from our point of view, no further regulatory actions are required.Return of product: we will return the investigated product to the sender for our own relief.
 
Event or Problem Description
It was reported that a shuntassistant (#fv250t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt showed an over-drainage and adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2023.The complaint device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 2 years, 7 months, weight: 14 kilograms (kg), height: 93.5 centimeters (cm), gender: male.Same event as #3004721439-2023-00109.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNTASSISTANT 15
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key17401695
Report Number3004721439-2023-00205
Device Sequence Number2494020
Product Code JXG
UDI-Device Identifier04041906136987
UDI-Public4041906136987
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110206
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2023
Device Model NumberFV250T
Device Catalogue NumberFV250T
Device Lot Number20037938
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 MO
Patient SexMale
Patient Weight14 KG
-
-