| Model Number |
T200-020 |
| Medical Device Problem Codes |
Melted (1385); Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All components of the device were tested prior to release.No issues were found.After shipment of the device to the customer, no shipping damage was reported.It is possible that the power cord was somehow damaged after delivery and unpackaging, prior to use by the customer.The device has not been returned yet.A follow-up report shall be filed when new information is available.There was no patient involvement or impact on the user.
|
| |
|
Event or Problem Description
|
|
A facility reported that upon setting up the device, with the power cord plugged into the wall outlet and into the power supply, the power cord started to get warm and generate smoke.
|
| |
|