| Catalog Number |
11101-000021 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that the device would not detect the electrodes being connected.As a result, defibrillation therapy would not be available, if needed.There were no reports of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was unable to duplicate the issue of the electrodes not connecting correctly with the device.Upon inspection, it was observed that the device kept resetting after the lid is opened.The device can no longer be repaired.The customer was provided with a replacement device.The cause of the reported issue could not be determined.
|
| |
|
Event or Problem Description
|
|
A customer contacted stryker to report that the device would not detect the electrodes being connected.As a result, defibrillation therapy would not be available, if needed.There were no reports of patient use associated with the reported event.
|
| |
|