It was reported by netherlands that during service and evaluation, it was determined that the attachment device generated heat.It was further observed that the device had a worn gear, and the cutter device could not be inserted.It was further determined that the device failed pretest for cutter insertion and temperature.It was noted in the service order that the device was not functioning properly.It was further reported that the drill device had difficulty sliding though the attachment.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese, or its employees that the report constitutes an admission that the product, depuy synthese, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: drill device, 1/1/2023.Device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the drill device having difficulty sliding through the attachment device was not confirmed.Therefore, an assignable root cause was not determined.However, the device generating heat, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|