• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number MA-15C
Medical Device Problem Codes Vibration (1674); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: h6: device history review: a device history review was performed which indicated that there were no relevant issues identified during the manufacture of the device that may have contributed to the reported condition.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device produced vibration was not confirmed.Therefore, an assignable root cause for the reported condition of vibration was not determined.However, during evaluation, it was determined that the device produced heat.The assignable root cause was determined to be due to component failure from wear.Udi: (b)(4).
 
Event or Problem Description
It was reported by switzerland that during service and evaluation, it was determined that the curved bearing sleeve device had a worn bearing, damaged components, produced heat and had component damage.It was further determined that the device failed pretest for visual assessment and temperature assessment.It was noted in the service order that the device produced vibration.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2023.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BEARING SLEEVE, 15CM CURVED
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 
3035526892
MDR Report Key17405689
Report Number1045834-2023-01136
Device Sequence Number12602118
Product Code HBE
UDI-Device Identifier00845384004325
UDI-Public00845384004325
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K042783
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMA-15C
Device Lot NumberL033121280
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Initial Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-