| Model Number |
232-10-85-300 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The threads on the proximal end of the graft bolt that latch into the cage reportedly sheared off.The graft bolt went straight through the cage and wouldn't lock.The surgeon removed the graft bolt and implanted a screw that locked into the cage.
|
| |
|
Additional Manufacturer Narrative
|
|
The graft bolt has not returned for evaluation.Photographs were not provided.The identifying lot number was not provided; therefore, a review of the device history records could not be performed.The root cause could not be determined at this time.If additional information is received, a supplemental report will be submitted.
|
| |
|