| Catalog Number |
201038 |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility reported one of the 10 packs of unopened scalp clip disposable (id 201038) has a brown dot.The issue was detected before the surgery.The procedure was completed with a replacement product available.No patient injury reported, and the event did not led to surgical delay.Based on the information provided, it is unknown if the product was stored in its original packaging, in its original pouch, or original dispenser box.It is also unknown how was the product stored.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Scalp clip disposable (id 201038) was returned for evaluation.Device history record (dhr) - no anomalies occurred during the manufacturing process.Failure analysis - in the failure analysis that was performed, the returned unit was found to have the brown dot.Root cause analysis - the likely root cause is a residue present at the time of molding the parts of the applier.
|
| |
|