MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. TMC 3.0MM COMPRESSION SCREW SYSTEM - INTERFRAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. TMC 3.0MM COMPRESSION SCREW SYSTEM - INTERFRAG Back to Search Results
Model Number SK18
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2022, all hardware was removed in a revision surgery on (b)(6) 2023 due to nonunion, a broken plate and broken screw.The tip of the broken screw was retained in the patient's bone.The site was revised with non-tmc hardware and no other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records were reviewed, and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.Although a number of factors could have contributed to what was reported, the most likely cause cannot be determined due to the device not being returned for evaluation.However, it is possible the devices were subjected to excessive bending stresses leading to their breakage.The company will supplement this mdr as necessary and appropriate.The broken plate was reported in mdr 3011623994-2023-00186.
 
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2022, all hardware was removed in a revision surgery on (b)(6) 2023 due to nonunion, a broken plate and broken screw.The tip of the broken screw was retained in the patient's bone.The site was revised with non-tmc hardware and no other patient outcomes were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMC 3.0MM COMPRESSION SCREW SYSTEM - INTERFRAG
Common Device Name
SCREW
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
uriza shums
100 palmetto park place
ponte vedra, FL 32081
9043735940
MDR Report Key17407164
Report Number3011623994-2023-00187
Device Sequence Number6093438
Product Code HWC
UDI-Device Identifier00853114006068
UDI-Public00853114006068
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K213036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/24/2025
Device Model NumberSK18
Device Lot Number300101425
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2023
Initial Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-