| Catalog Number |
TRM-445 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported a left side implant exchange due to concerns with the product.Healthcare professional later reported rupture detected during explant surgery.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device exchange.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: anxiety-product/procedure: unable to observe since it is not related to the device.Rupture: observed a broken-on posterior assessed as an unidentified (tear) opening (shell thickness within spec) additional observations: wear abrasion observed on the device.No further actions are required as the device was implanted.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side implant exchange due to concerns with the product.Healthcare professional later reported rupture detected during explant surgery.Device has been explanted.
|
| |
|