| Catalog Number |
P/N 397001-001 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The companion driver caddy will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
|
| |
|
Event or Problem Description
|
|
The customer, a syncardia certified hospital, reported that a c2 driver was stuck in the caddy - they can't press in the lock button to remove the driver.States driver is seated properly in the caddy groove.Email later received that biomed was finally able to remove the driver from the caddy, but they are requesting to send it back to be checked out.
|
| |
|
Additional Manufacturer Narrative
|
|
Device history record (dhr) review confirmed that companion caddy s/n ca09000340 was serviced and passed all functional teting prior to being released to finished goods.Visual inspection of internal and external components found no abnormalities.Companion caddy passed all functional testing for acceptance at incoming inspection.An addtiional driver docking test was performed with the driver docked and then undocked ten times.No fault or issue was observed during this testing.Complaint was unable to be replicated.Failure investigation for this complaint confirmed the reported issue from the photos provided with the initial complaint.The customer complaint was not replicated during testing; the root cuase of the reported malfunction of the caddy to undock the driver was unable to be determined.Failure investigation identified no test failures or damage that could have contributed to the complaint.No evidence of a device malfunction was found.Device was not in patient use at time of complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.
|
| |
|
Event or Problem Description
|
|
The customer, a (b)(6) hospital, reported that a c2 driver was stuck in the caddy - they can't press in the lock button to remove the driver.States driver is seated properly in the caddy groove.Email later received that biomed was finally able to remove the driver from the caddy, but they are requesting to send it back to be checked out.
|
| |
|