• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI; FORCEPS, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI; FORCEPS, LAPAROSCOPIC Back to Search Results
Model Number WA64150A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that the part that opens and closes the forceps was damaged.It is presumed that the damage was caused by a load exceeding the resistance of the damaged part.There were no missing parts.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
The customer reported to olympus that during a therapeutic, laparoscopic colectomy procedure, the root portion of the grip is deformed and detached from the jaws 5 x 330 mm, grasping forceps.The procedure was completed using a similar device.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported root portion of the grip detachment from the jaws was found to be the result of a broken pull rod at the distal end.The broken pull rod was likely caused by excessive force exerted to the jaw during user handling/mishandling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI
Common Device Name
FORCEPS, LAPAROSCOPIC
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17409644
Report Number9610773-2023-02027
Device Sequence Number14927662
Product Code HET
UDI-Device Identifier04042761068970
UDI-Public04042761068970
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K944201
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64150A
Device Lot Number21Z01-007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2023
Supplement Date Received by Manufacturer03/26/2024
Initial Report FDA Received Date07/27/2023
Supplement Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A60800A A60201A
Patient SexUnknown
-
-