| Model Number |
WA64150A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that the part that opens and closes the forceps was damaged.It is presumed that the damage was caused by a load exceeding the resistance of the damaged part.There were no missing parts.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that during a therapeutic, laparoscopic colectomy procedure, the root portion of the grip is deformed and detached from the jaws 5 x 330 mm, grasping forceps.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported root portion of the grip detachment from the jaws was found to be the result of a broken pull rod at the distal end.The broken pull rod was likely caused by excessive force exerted to the jaw during user handling/mishandling.Olympus will continue to monitor field performance for this device.
|
| |
|