| Model Number |
MS18853 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The hospital technician found the low-voltage differential signaling (lvds) connection cable was loose, reconnecting the cable resolved the issue.No device malfunction was identified.The device was returned to use.H3 other text : the device was evaluated on site with no malfunctions identified.
|
| |
|
Event or Problem Description
|
|
It was reported that when the anesthesiology department used the monitor to monitor the vital signs of patients undergoing ophthalmic surgery, the screen suddenly went black.After the equipment maintenance engineer arrived, he used an external display screen for the operation to continue.Though no adverse patient impact was reported, a delay in treatment cannot be excluded for having occurred.
|
| |
|