EXACTECH, INC. LOGIC CR TIB INSERT STD, SZ 2, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
02-012-47-2009 |
| Medical Device Problem Codes |
Naturally Worn (2988); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Discomfort (2330); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had an initial right tka on (b)(6) 2018.The patient was revised on (b)(6) 2023 due to the poly recall.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.H7: z-0021-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: corrected the following: health effect - clinical code, medical device problem code, component code, type of investigation, investigation findings, investigation conclusions manufacturer narrative updated: the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear could not be confirmed and potential contributions of user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images, radiographs, or relevant clinical information was provided.
|
| |
|