• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD Back to Search Results
Model Number 2140002
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient previously implanted with plates and screws in a 2 level olif for degenerative disc disease at l2/3 & l3/4.It was reported that the surgeon noticed that the plates and screws had backed out of the vertebral bodies during a follow up appointment with the patient.The patient had ongoing back pain.The surgeon is planning posterior fixation at these levels.The date of initial operation was (b)(6) 2023.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
B2: other- posterior fixation planned h6: the device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Additional event details medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A posterior fixation and decompression of l2-l4 was performed on (b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIVOX¿ OBLIQUE LATERAL SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key17411901
Report Number1030489-2023-00540
Device Sequence Number20619038
Product Code KWQ
UDI-Device Identifier00643169594234
UDI-Public00643169594234
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K152277
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2140002
Device Catalogue Number2140002
Device Lot Number0737828W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2023
Supplement Date Received by Manufacturer03/08/2024
Initial Report FDA Received Date07/28/2023
Supplement Report FDA Received Date03/13/2024
Date Device Manufactured09/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-