MEDOS INTERNATIONAL SàRL CH VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 7 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION Back to Search Results
Catalog Number 186727745
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/04/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in austria as follows: it was reported that the patient underwent a revision surgery on (b)(6) 2023 due to a broken expedium cfx screw that was implanted in s1 as a part of a long-distance degenerative treatment.During the surgery, it was determined that further screws had cooed into the pedicles.These had to be replaced with new expedium cfx implants.When removing the loosened implants with the expedium polydriver, the screw heads broke off.When implanting the revision screws, an expedium polydriver broke off at the drive of the screwdriver in the shaft head of the screw despite proper handling.Additionally, two awls were bent.This report involves one viper system fenestrated cortical fix polyaxial screw 5.5 x 7 x 45mm.This is report 2 of 7 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.D2b: additional device product codes: pml.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device visual analysis of the photo revealed that the [mis ti cfx fen poly 7x45] had broken approximately between the shaft and head junction.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the [mis ti cfx fen poly 7x45].There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
1st/2nd: (b)(6) 2022, implantation of the cfx/ilium screws.A loose screwdriver attachment that remained in the s1 screw, during the procedure on (b)(6) 2022.For revision, procedure was completed successfully with no surgical delay.Non-union at l5/s1 was reported.No fragments were left behind in the patient, due to broken screws.Intervention was required to remove all fragments.Patient status was noted, as a decrease of pain medication.And a discharge to contemporary nursery home.No further follow up for this complaint, we will be reporting the post-operative and intra-operative event on (b)(4).The reported, screws had malfunctions, during post-operative and intra-operative phases of care.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 X 7 X 45MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key17412406
Report Number1526439-2023-01377
Device Sequence Number6983192
Product Code MNI
UDI-Device Identifier10705034282671
UDI-Public(01)10705034282671
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K160879
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number186727745
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/06/2023
Supplement Date Received by Manufacturer08/14/2023
09/08/2023
Initial Report FDA Received Date07/28/2023
Supplement Report FDA Received Date08/28/2023
09/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MIS TI CFX FEN POLY 7X45; MIS TI CFX FEN POLY 7X45; MIS TI CFX FEN POLY 7X45; MIS TI CFX FEN POLY 7X50; MIS TI CFX FEN POLY 7X50; XPDM PEDICLE PROBE, STRAIGHT; XPDM QUICK-CON SI POLY SCWDRVR; XPDM THORACIC PEDICLE PRB, ST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight76 KG
-
-