| Catalog Number |
BRZ-AP1-080-000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hyperplasia (1906)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This is a historical record and reflects reports that were received by the company prior to april 2021.This mdr is associated with a parallel plate applicator, these applicators were discontinued and recalled in 2022.According to the coolsculpting user manual, under rare adverse events, paradoxical adipose hyperplasia (ph/pah) is characterized by a visibly enlarged tissue volume within the treatment area, which may develop two to five months after treatment.Surgical intervention may be required.Pah is not related to any coolsculpting device failure mode but it is observed more frequently with the parallel plate applicators.Pah is included in the risk management files of the device as an inherent risk to the use of cryolipolysis for localized fat reduction.
|
| |
|
Event or Problem Description
|
|
Allergan aesthetics received a report of a female patient who was treated with coolsculpting on (b)(6) 2012 with ez app 8.0 applicator to the lower abdomen.The patient was assessed by a physician who diagnosed the lower abdomen with paradoxical hyperplasia.
|
| |
|
Event or Problem Description
|
|
Allergan aesthetics received a report of a female patient who was treated with coolsculpting on (b)(6) 2012 with ez app 8.0 applicator to the lower abdomen.The patient was assessed by a physician who diagnosed the lower abdomen with paradoxical hyperplasia.
|
| |
|
Search Alerts/Recalls
|