| Model Number |
799-EMVP-05 |
| Medical Device Problem Codes |
Calibration Problem (2890); Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed "runtime calibration failure - 1051" and "5v self check failure - 1172" error messages.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including functional testing, calibration testing, exhalation backup and stress testing without duplicating the report.The smart pneumatic module (spm) board and flow screens were replaced as a pre-caution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|