|
It was reported that on (b)(6) 2023, while drilling for with a 2.7 cannulated drill bit, the wire became incarcerated inside of the drill bit and was pulled out.It appeared that the surgeon was drilling at an angle and it caused the wire to bind up inside of drill bit.Another drill bit and wire were selected and the surgery was successfully completed.There was 5 minutes surgical delay.Procedure was completed successfully.Patient outcome is stable.This report is for one 1.25mm non-threaded guide wire 150mm for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|